Data Foundry

Medical device recalls 2025

Corin Operating Tables Model Numbers: — Class II medical device recall Z-2420-2025

Ongoing recall by MAQUET GMBH, reported 2025-09-03, distributed nationwide. Under certain conditions, such as the simultaneous use of the operating table with other medical devices, incl

Recall numberZ-2420-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMAQUET GMBH, Rastatt, Germany
Recall initiated2025-07-14
Classified2025-08-22
Reported by FDA2025-09-03
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BG, CA, CH, CN, CO, DE, DK, DZ, EE, EG, ES, FI, FR, GB, IE, IL, IT, JO, JP, KR, KW, LB, MA, NL, NO, PL, PT, RO, SA, SE, SG, TH, TR, ZA
UPC / GTIN / UDI-DI04046768145177, 04046768145184, 04046768145191, 04046768145207, 04046768145214, 04046768145221
Serial numbers100, 101, 102, 103, 104, 105, 106, 107, 108, 109, 110, 111, 112, 113, 114, 115, 116, 118, 119, 120, 121, 122, 123, 124, 125 and 39 more
Model numbers04046768145177, 04046768145184, 04046768145191, 04046768145207, 04046768145214, 04046768145221, 100, 101, 102, 103, 104, 105, 106, 107, 108, 109, 110, 111, 112, 113, 114, 115, 116, 118, 119 and 51 more

Product

Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.

Reason for recall

Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2420-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.