Corin Operating Tables Model Numbers: — Class II medical device recall Z-2420-2025
Ongoing recall by MAQUET GMBH, reported 2025-09-03, distributed nationwide. Under certain conditions, such as the simultaneous use of the operating table with other medical devices, incl
| Recall number | Z-2420-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | MAQUET GMBH, Rastatt, Germany |
| Recall initiated | 2025-07-14 |
| Classified | 2025-08-22 |
| Reported by FDA | 2025-09-03 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BG, CA, CH, CN, CO, DE, DK, DZ, EE, EG, ES, FI, FR, GB, IE, IL, IT, JO, JP, KR, KW, LB, MA, NL, NO, PL, PT, RO, SA, SE, SG, TH, TR, ZA |
| UPC / GTIN / UDI-DI | 04046768145177, 04046768145184, 04046768145191, 04046768145207, 04046768145214, 04046768145221 |
| Serial numbers | 100, 101, 102, 103, 104, 105, 106, 107, 108, 109, 110, 111, 112, 113, 114, 115, 116, 118, 119, 120, 121, 122, 123, 124, 125 and 39 more |
| Model numbers | 04046768145177, 04046768145184, 04046768145191, 04046768145207, 04046768145214, 04046768145221, 100, 101, 102, 103, 104, 105, 106, 107, 108, 109, 110, 111, 112, 113, 114, 115, 116, 118, 119 and 51 more |
Product
Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.
Reason for recall
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.