Data Foundry

Medical device recalls 2018

Hyalomatrix — Class II medical device recall Z-2423-2018

Terminated recall by Anika Therapeutics, Inc., reported 2018-07-18, distributed nationwide. Hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging m

Recall numberZ-2423-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAnika Therapeutics, Inc., Bedford, MA, United States
Recall initiated2018-05-10
Classified2018-07-11
Reported by FDA2018-07-18
Terminated2019-05-10
DistributedNationwide (US)
Quantity1,711
Reason classesspecification failure

Product

Hyalomatrix, Product nos. 651105, 651103, 651104, and 651301 Product Usage: Hyalomatrix is a bi-layered, sterile, flexible, and comfortable wound dressing that acts as an advanced wound care device. It is comprised of a non-woven pad entirely composed of HY AFF 11, a benzyl ester of hyaluronic acid, and a semipermeable silicone membrane, which controls water vapor loss, provides a flexible covering for the wound surface, and adds increased tear strength to the device. The HYAFFl 1 wound contact layer biodegradable matrix acts as a scaffold for cellular invasion and capillary growth. Hyalomatrix is indicated for the management of wounds including: partial and full-thickness wounds; second-degree bums; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use.

Reason for recall

Hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging material during sterilization, which resulted in degradation of the film. Over the long term it is possible that product integrity may be compromised.