STERIS 5085 and 5085 SRT Surgical Tables — Class II medical device recall Z-2426-2012
Terminated recall by Steris Corporation, reported 2012-09-26, distributed nationwide. STERIS has learned from Customer feedback and field service experience that damage from external impact to the
| Recall number | Z-2426-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Steris Corporation, Mentor, OH, United States |
| Recall initiated | 2012-04-06 |
| Classified | 2012-09-20 |
| Reported by FDA | 2012-09-26 |
| Terminated | 2013-09-04 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, CN, IN, KR, MX, PA, TH |
| Quantity | 379 tables |
Product
STERIS 5085 and 5085 SRT Surgical Tables, Product Usage: The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. It accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients. The 5085SRT table is a variation of the 5085 table that enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery).
Reason for recall
STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the stainless steel shroud sections covering the table hydraulic column to become misaligned; if the damage is severe, the up/down movement of the table may be impeded. In addition, the table hand control may indicate that the table floor locks are unlock