Data Foundry

Medical device recalls 2024

ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX… — Class II medical device recall Z-2426-2024

Ongoing recall by Radiometer Medical ApS, reported 2024-07-31, distributed nationwide. Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially bec

Recall numberZ-2426-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmRadiometer Medical ApS, Bronshoj, Denmark
Recall initiated2024-05-31
Classified2024-07-24
Reported by FDA2024-07-31
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA
Quantity27,127 analyzers
Reason classesmicrobial contamination
UPC / GTIN / UDI-DI05700693938004, 05700693938219, 05700693938264
Model numbers393-800, 393-801, ABL800

Product

ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) " In vitro testing of samples of expired air for the parameters pO2 and pCO2 " In vitro testing of pleura samples for the pH parameter.

Reason for recall

Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2426-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.