WATCHMAN FXD Curve Access Sys Sgl — Class I medical device recall Z-2429-2025

Ongoing recall by Boston Scientific Corporation, reported 2025-09-10. Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed with

Recall numberZ-2429-2025
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Maple Grove, MN, United States
Recall initiated2025-07-29
Classified2025-08-29
Reported by FDA2025-09-10
Quantity340,185 units
UPC / GTIN / UDI-DI00191506013820

Product

WATCHMAN FXD Curve Access Sys Sgl, OUS, Material Number (UPN) M635TS80010; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Reason for recall

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.