Data Foundry

Medical device recalls 2023

Cardiosave Hybrid Intra-Aortic Balloon Pumps — Class I medical device recall Z-2431-2023

Ongoing recall by Datascope Corp., reported 2023-09-06, distributed nationwide. Users were reporting that the device was not charging as expected. It was discovered that users were unaware t

Recall numberZ-2431-2023
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmDatascope Corp., Mahwah, NJ, United States
Recall initiated2023-07-31
Classified2023-08-30
Reported by FDA2023-09-06
DistributedNationwide (US)
Quantity9,175 units
UPC / GTIN / UDI-DI10607567108391, 10607567108407, 10607567108414, 10607567108421, 10607567108438, 10607567109008, 10607567109053, 10607567109107, 10607567111117, 10607567111940, 10607567112312, 10607567113432, 10607567113449
Model numbers0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55

Product

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65

Reason for recall

Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the cart the battery(ies) will not charge.