Siemens Symbia T Series — Class II medical device recall Z-2436-2015

Terminated recall by Siemens Medical Solutions USA, Inc., reported 2015-08-26, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo syst

Recall numberZ-2436-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc., Hoffman Estates, IL, United States
Recall initiated2015-07-31
Classified2015-08-20
Reported by FDA2015-08-26
Terminated2016-12-23
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, HK, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KR, KW, LT, LU, LV, MA, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, VE, VN, ZA
Quantity1,700
Reason classesdevice malfunction
Model numbers10275007, 10275008, 10275009, 10275010, 8717733

Product

Siemens Symbia T Series, a radiological imaging system that is a combination of a single photon emission computed tomography (SPECT) camera system for nuclear medicine images, and a computed tomography (CT) camera system for x-ray images.

Reason for recall

Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to the overheating of the assembly.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2436-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.