Siemens Intevo Series system — Class II medical device recall Z-2437-2015
Terminated recall by Siemens Medical Solutions USA, Inc., reported 2015-08-26, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo syst
| Recall number | Z-2437-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Medical Solutions USA, Inc., Hoffman Estates, IL, United States |
| Recall initiated | 2015-07-31 |
| Classified | 2015-08-20 |
| Reported by FDA | 2015-08-26 |
| Terminated | 2016-12-23 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, HK, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KR, KW, LT, LU, LV, MA, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, VE, VN, ZA |
| Quantity | 185 |
| Reason classes | device malfunction |
| Model numbers | 10764801, 10764802, 10764803, 10764804 |
Product
Siemens Intevo Series system, a radiological imaging system that is a combination of a single photon emission computed tomography (SPECT) camera system for nuclear medicine images, and a computed tomography (CT) camera system for x-ray images.
Reason for recall
Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to the overheating of the assembly.