Data Foundry

Medical device recalls 2020

(1) Atellica Data Manager software version 1.0 Siemens Material… — Class II medical device recall Z-2438-2020

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2020-07-08, distributed nationwide. Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Numbe

Recall numberZ-2438-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States
Recall initiated2020-05-15
Classified2020-06-26
Reported by FDA2020-07-08
Terminated2021-05-18
DistributedNationwide (US)
Countries outside the USAE, AR, AU, BE, BR, BS, CL, CO, CZ, DE, EG, ES, GB, GR, HU, IN, IT, KW, LT, LV, MX, MY, OM, PE, PL, PT, QA, SA, SE, SG, TH, TR, ZA
Quantity70 units
UPC / GTIN / UDI-DI00630414234946

Product

(1) Atellica Data Manager software version 1.0 Siemens Material Number (SMN): 11314237 (2) Atellica Data Manager software version 1.1 Siemens Material Number (SMN): 11316888 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test ordering and priority routing information to automation, apply rules to test orders and auto-verification rules for test results, and provide integrated management of quality control of test system.

Reason for recall

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2438-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.