Data Foundry

Medical device recalls 2012

BP7561 OPTI CCA-TS B-Lac Cassette — Class II medical device recall Z-2440-2012

Terminated recall by OPTI Medical Systems, Inc, reported 2012-10-03, distributed nationwide. The values reported for PCO2 by BP7561 B-Lac cassettes are biased at high concentrations when used at altitude

Recall numberZ-2440-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOPTI Medical Systems, Inc, Roswell, GA, United States
Recall initiated2012-07-03
Classified2012-09-21
Reported by FDA2012-10-03
Terminated2013-04-11
DistributedNationwide (US)
Countries outside the USAR, BO, CH, CN, DE, EC, FR, GB, GR, ID, IE, IN, IR, JP, KW, LT, LU, MA, MX, MY, NG, PL, RO, RU, SA, SG, TR, VE, VN, ZA
Lot numbers110650, 114650, 114651, 114652, 114653, 116650, 116652, 117650, 118651, 121650, 121651, 122652, 122653, 122654, 122655, 122656, 122657, 123654, 123655, 124650, 133650, 134651, 136650, 146650, 149650, 202650, 208650, 208651, 209560, 209651, 211650, 212651, 213650, 214650, 215650, 215651, 221650, 222650, 224650

Product

BP7561 OPTI CCA-TS B-Lac Cassette (25 per box) Cassettes are consumables used in the OPTI CCA-TS Analyzer intended to be used for in vitro measurements of pH, PCOX2 PO2, lactate (lactic acid), total hemoglobin (tHb), and oxygen saturation (SO2), in heparinized whole blood samples (either arterial or venous) on the OPTI CCA-TS system, in either a clinical setting or point of care locations. Cassettes contain one-time use sensors that are used with the OPTI CCA-TS analyzer to perform in-vitro measurements of blood gases and metabolites. OPTI CCA-TS B-Lac Cassettes are supplied in boxes of 25 and may only used with the OPTI CCA-TS Analyzer.

Reason for recall

The values reported for PCO2 by BP7561 B-Lac cassettes are biased at high concentrations when used at altitudes above 2800 feet (853 meters).