Data Foundry

Medical device recalls 2015

iChemVELOCITY Automated Urine Chemistry Catalog No — Class II medical device recall Z-2441-2015

Terminated recall by Iris Diagnostics, reported 2015-09-02, distributed nationwide. Iris International is recalling the iChemVELOCITY systems because they do not maintain the on-board strip stab

Recall numberZ-2441-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIris Diagnostics, Chatsworth, CA, United States
Recall initiated2015-07-15
Classified2015-08-21
Reported by FDA2015-09-02
Terminated2016-05-26
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BH, BO, BR, CA, CH, CL, CN, CO, CZ, DE, EC, ES, FR, GB, GR, HK, HR, HU, IL, IN, IT, JO, KR, KW, LB, LU, MA, MX, MY, NL, NZ, OM, PA, PE, PH, PL, PT, PY, QA, RO, RS, SA, SG, SK, TH, TR, TW, VE, ZA
Quantity2,448 units
Reason classeslabeling, specification failure
Model numbers800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713, 800-7714, 800-7719, 800-7720

Product

iChemVELOCITY Automated Urine Chemistry Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713,800-7714, 800-7719, 800-7720. An in vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips.

Reason for recall

Iris International is recalling the iChemVELOCITY systems because they do not maintain the on-board strip stability claim of 5 days at 18-28 degrees celsius at 20-80% relative humidity (RH) as stated in the iChemVELOCITY product labeling.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2441-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.