Data Foundry

Medical device recalls 2012

SharePlan 1 — Class II medical device recall Z-2444-2012

Terminated recall by RAYSEARCH LABORATORIES AB, reported 2012-10-03, distributed in CA, KY, MI, NY, OK, OR, SD, WI. This notice concerns behavior of the SharePlan that may be unexpected and not clearly described in the user ma

Recall numberZ-2444-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRAYSEARCH LABORATORIES AB, Stockholm, N/A, Sweden
Recall initiated2011-10-10
Classified2012-09-24
Reported by FDA2012-10-03
Terminated2013-01-25
DistributedCA, KY, MI, NY, OK, OR, SD, WI
Countries outside the USAU, BE, CA, DE, FR, GB, PL, SE
Quantity20 units

Product

SharePlan 1.1, aka t-RayAutoplan version 1.2.1 (build number 1.3.1.10). Manufacturing name RayAutoplan, aka t-RayAutoplan, commercial name (TomoTherapy) SharePlan Intended use: t-RayAutoplan is a software that, based on the planned dose distribution from a TomoPlan treatment planning system, generates a family of individually optimized IMRT plans and presents their characteristics to the user in a GUI. Among the generated plans, the user selects and clinically approves a treatment plan and exports it electronically to DICOM, for subsequent treatment of the patient. The intended users of t-RayAutoplan shall be clinically qualified staff, such as medical physicists, medical doctors or dosimetrists.

Reason for recall

This notice concerns behavior of the SharePlan that may be unexpected and not clearly described in the user manual. This behavior has not caused patient mistreatment or other incidents. However, the described user actions must be adhered to for best agreement between treatment plan evaluation and the clinical outcome of the delivered treatment. The behavior appears, when importing RFA-files (.asc)