Data Foundry

Medical device recalls 2015

cobas b 123 Fluid Pack COOX REF 05169992001 200 Product Usage: The… — Class II medical device recall Z-2446-2015

Terminated recall by Roche Diagnostics Operations, Inc., reported 2015-09-02, distributed nationwide. Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration

Recall numberZ-2446-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Operations, Inc., Indianapolis, IN, United States
Recall initiated2015-07-14
Classified2015-08-21
Reported by FDA2015-09-02
Terminated2016-01-19
DistributedNationwide (US)
Quantity82 fluid
Lot numbers21446113
Model numbers05169992001, 200
Expiration dates2015-08-08, 2015-08-18, 2015-09-18, 2015-09-19, 2015-10-01

Product

cobas b 123 Fluid Pack COOX REF 05169992001 200 Product Usage: The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes NAT, K+, iCaWt (ISE), hematocrit (THct), metabolites (GLu, Lac) total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COHb, MetHb), and oxygen saturation (SO2). The cobas b 123 POC system calculates derived parameters. Depending on the equipment configuration of the instrument, the Sensor Cartridge and the Fluid Pack used, the following parameters are measured in human whole blood and QC materials.

Reason for recall

Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits. Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2446-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.