Data Foundry

Medical device recalls 2023

The Air/Water Valve is provided/used with the following Endoscopes:… — Class II medical device recall Z-2447-2023

Ongoing recall by Olympus Corporation of the Americas, reported 2023-08-30, distributed nationwide. The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-

Recall numberZ-2447-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOlympus Corporation of the Americas, Center Valley, PA, United States
Recall initiated2023-07-12
Classified2023-08-23
Reported by FDA2023-08-30
DistributedNationwide (US)
Quantity29,590 units
UPC / GTIN / UDI-DI04953170355929

Product

The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASONIC GASTROVIDEOSCOPE, ULTRASONIC COLONOVIDEOSCOPE: GF-UC140P-AL5, GF-UCT140-AL5, GF-UE160-AL5, GF-UCT180, GF-UM20, GF-UM130, GF-UMQ130, GF-UM160, GF-UC160P-OL5, GF-UCT160-OL5

Reason for recall

The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body fluid could backflow into the air/water channel of the ultrasonic endoscopes during the procedure.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2447-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.