Data Foundry

Medical device recalls 2023

ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein — Class II medical device recall Z-2451-2023

Ongoing recall by Siemens Healthcare Diagnostics, Inc., reported 2023-08-30, distributed nationwide. There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzym

Recall numberZ-2451-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States
Recall initiated2023-07-25
Classified2023-08-24
Reported by FDA2023-08-30
DistributedNationwide (US)
Quantity3,067 units
Reason classespotency
UPC / GTIN / UDI-DI00630414279176

Product

ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid

Reason for recall

There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.