ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein — Class II medical device recall Z-2451-2023
Ongoing recall by Siemens Healthcare Diagnostics, Inc., reported 2023-08-30, distributed nationwide. There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzym
| Recall number | Z-2451-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States |
| Recall initiated | 2023-07-25 |
| Classified | 2023-08-24 |
| Reported by FDA | 2023-08-30 |
| Distributed | Nationwide (US) |
| Quantity | 3,067 units |
| Reason classes | potency |
| UPC / GTIN / UDI-DI | 00630414279176 |
Product
ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid
Reason for recall
There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.