BD Alaris Pump Infusion Buerette Set 0 — Class II medical device recall Z-2459-2021
Ongoing recall by CAREFUSION, reported 2021-09-22, distributed nationwide. Needle-free connector valves, either standalone or attached to extension sets, may have issues including diffi
| Recall number | Z-2459-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | CAREFUSION, San Diego, CA, United States |
| Recall initiated | 2021-08-02 |
| Classified | 2021-09-13 |
| Reported by FDA | 2021-09-22 |
| Distributed | Nationwide (US) |
| Quantity | 344,300 |
| Reason classes | device malfunction |
| Lot numbers | 20015346, 20015347, 20016918, 20016919, 20016960, 20016961, 20017112, 20017400, 20017401, 20017423, 20025114, 20025123, 20026226, 20026227, 20026322, 20026323, 20026382, 20026383, 20026510, 20035102, 20035133, 20035134, 20035281, 20035295, 20035296, 20035368, 20035526, 20035527, 20035603, 20035604, 20036962, 20055717, 20056757, 20056758, 20056830, 20056850, 20056851, 20056946, 20056947, 20057055 and 118 more |
| Model numbers | 10015012, 10817920, 10821753, 11607787, 11613191, 2440-0600, 2441-0007, 2444-0007, 2447-0007, C24104E |
Product
BD Alaris Pump Infusion Buerette Set 0.2 Micron Filter Smallbore Tubing SmartSite Port (Burette) SmartSite Y-Site, REF: 10015012; BD Alaris Pump Infusion Buerette Set Ball Valve Smallbore Tubing Smartsite Port (Burette) SmartSite Y-Site, REF: 10817920; BD Alaris Pump Infusion Buerette Set SmartSite Port (Burette) 3 SmartSite Y-Sites, REF: 10821753; BD Alaris Pump Infusion Buerette Smallbore Tubing Smartsite Port (Burette) SmartSite Y-Site, REF: 11607787; BD Alaris Pump Infusion Buerette Set 0.2Micron Filter Ball Valve SmartSite Port (Burette) 2 SmartSite Y-Sites, REF: 11613191; BD Alaris Pump Infusion Buerette Set Smartsite Port (Burette) 3 SmartSite Y-Sites, REF: 2440-0600; BD Alaris Pump Infusion Buerette Set Smallbore Tubing Smartsite Port (Burette) 3 SmartSite Y-Site, REF: 2441-0007; BD Alaris Pump Infusion Buerette Set Low Sorbing Tubing Smartsite Port (Burette), REF: 2444-0007; BD Alaris Pump Infusion Buerette Set Ball Valve Smartsite Port (Burette) 2 Smartsite Y-Sites, REF: 2447-0007; BD Alaris Pump Infusion Set Back Check Valve 4-Way Stopcock 3 Smartsite Y-Sites Extension Set With SmartSite Port, REF: C24104E
Reason for recall
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2459-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.