Catheter Irrigation Set labeled in part: ***24K Arthroscopy… — Class II medical device recall Z-2469-2012
Terminated recall by Linvatec Corp. dba ConMed Linvatec, reported 2012-10-03, distributed nationwide. ConMed Linvatec recalled their 24k Arthroscopy Outflow/Suction Tubing Set (Product Number 24k100)because there
| Recall number | Z-2469-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Linvatec Corp. dba ConMed Linvatec, Largo, FL, United States |
| Recall initiated | 2011-07-01 |
| Classified | 2012-09-27 |
| Reported by FDA | 2012-10-03 |
| Terminated | 2013-07-25 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, CH, CN, CZ, DE, DK, EG, ES, FR, GB, IL, IT, NL, NZ, PL, PT, RU, SA, ZA |
| Quantity | 30,064 units |
| Reason classes | sterility, packaging |
| Lot numbers | 20853-000, 20854-000, 20855-000, 21791-000, 22172-000, 22414-000, 22513-000, 22667-000, 22724-000, 22725-000, 23606-000, 23912, 23912-000, 23971-000, 24177-000, 24296-000, 24348-000, 24373-000, 24476-000, 24559-000, 24671-000, 24703-000, 24905-000, 25668-000, 25794-000, 25931-000, 26025-000, 26041-000, 26294-000, 26448-000, 26486-000, 27130-000, 27255, 27255-000, 27322, 27520 |
| Model numbers | 24K100 |
Product
Catheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set*** This Arthroscopy Tubing Set is intended for use, in conjunction with the ConMed Linvatec 24k Pump, for pump-induced outflow and/or suction of fluid from the joint during arthroscopic procedures.
Reason for recall
ConMed Linvatec recalled their 24k Arthroscopy Outflow/Suction Tubing Set (Product Number 24k100)because there is a possibility that the product may have a breach in the Tyvek seal that could potentially compromise the sterility of the contents.