GS70 Salus Surgical Table — Class II medical device recall Z-2469-2023
Ongoing recall by Skytron, LLC, reported 2023-09-06, distributed nationwide. When the control of the operating table was used in Bluetooth mode, the operating table continued moving even
| Recall number | Z-2469-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Skytron, LLC, Grand Rapids, MI, United States |
| Recall initiated | 2023-08-01 |
| Classified | 2023-08-28 |
| Reported by FDA | 2023-09-06 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, NZ |
| Quantity | 110 units |
| Serial numbers | SUC-20207-0006, SUC-20207-0007, SUC-20207-0008, SUC-20207-0009, SUC-20207-0010, SUC-20207-0011, SUC-20207-0012, SUC-20207-0013, SUC-20208-0014, SUC-20208-0015, SUC-20208-0016, SUC-20208-0017, SUC-20208-0018, SUC-20208-0019, SUC-20208-0020, SUC-20208-0021, SUC-20208-0022, SUC-20208-0023, SUC-20208-0024, SUC-20208-0025, SUC-20208-0026, SUC-20208-0027, SUC-20208-0028, SUC-20208-0029, SUC-20209-0030 and 85 more |
Product
GS70 Salus Surgical Table, Product Code 18-070-70
Reason for recall
When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2469-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.