Data Foundry

Medical device recalls 2023

GS70H Salus Surgical Table — Class II medical device recall Z-2470-2023

Ongoing recall by Skytron, LLC, reported 2023-09-06, distributed nationwide. When the control of the operating table was used in Bluetooth mode, the operating table continued moving even

Recall numberZ-2470-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSkytron, LLC, Grand Rapids, MI, United States
Recall initiated2023-08-01
Classified2023-08-28
Reported by FDA2023-09-06
DistributedNationwide (US)
Countries outside the USAU, NZ
Quantity24 units
Serial numbersSUC-20207-0002, SUC-20207-0003, SUC-20208-0004, SUC-20208-0005, SUC-20208-0006, SUC-20208-0007, SUC-20209-0008, SUC-20209-0009, SUC-20209-0010, SUC-20209-0011, SUC-2020X-0012, SUC-2020X-0013, SUC-2020X-0014, SUC-2020X-0015, SUC-2020X-0016, SUC-2022Z-0017, SUC-2022Z-0018, SUC-2022Z-0019, SUC-2022Z-0020, SUC-2022Z-0021, SUC-20233-0022, SUC-20233-0023, SUC-20233-0024, SUC-20233-0025

Product

GS70H Salus Surgical Table, Product Code 18-071-70

Reason for recall

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.