Data Foundry

Medical device recalls 2020

uCT 530 Computed Tomography X-Ray System — Class II medical device recall Z-2472-2020

Ongoing recall by Shanghai United Imaging Healthcare Co., Ltd., reported 2020-07-08, distributed in TX. Two issues were identified with the computed tomography x-ray system including a service function which may ca

Recall numberZ-2472-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmShanghai United Imaging Healthcare Co., Ltd., Shanghai, N/A, China
Recall initiated2020-05-08
Classified2020-06-30
Reported by FDA2020-07-08
DistributedTX
Quantity4 units
UPC / GTIN / UDI-DI06971576831012

Product

uCT 530 Computed Tomography X-Ray System

Reason for recall

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2472-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.