Alinity ci-series System Control Module clinical chemistry and… — Class II medical device recall Z-2472-2021
Ongoing recall by Abbott Laboratories, reported 2021-09-22, distributed nationwide. Due to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There
| Recall number | Z-2472-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Laboratories, Irving, TX, United States |
| Recall initiated | 2021-07-26 |
| Classified | 2021-09-14 |
| Reported by FDA | 2021-09-22 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JM, JO, JP, KE, KR, KW, LB, LT, LV, MA, MX, MY, NI, NL, NO, NZ, OM, PE, PH, PK, PL, PT, PY, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TT, TW, UA, UY, VN, ZA |
| Quantity | 4,833 systems |
| Reason classes | specification failure, packaging, device malfunction, software |
| Model numbers | 03R70-01 |
Product
Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer for in vitro diagnostic use. Model number LN 03R70-01
Reason for recall
Due to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There is the potential for onboard reagents past their lot expiration date or onboard stability time to be incorrectly displayed on the Reagent Status screen with a status of "OK". In this scenario, the expired reagent will remain in the carousel for sample processing. 2) Performing monthly maintenance procedure 5701 Clean ICT Drain Tip incorrectly may potentially cause issues such as damaged connector or leaking connection. 3) A sample exception with message code 150 "Unable to process test. Previous processing module error" is produced without notifying the operator of a reagent pipettor failure and may cause processing module to transition into the Pausing state. Tests initiated prior to the sample exception may continue to process without dispense of reagents, potentially leading to incorrect results. 4. Reagent short dispense: Alinity c needs a control measure to protect against the potential for short reagent dispense. Addition of a maximum depth limit for reagent cartridges on the Alinity c to prevent the potential for a short dispense is needed. 5. RSM indicators were blinking green incorrectly: The Reagent and Sample Manager (RSM) indicators will incorrectly blink green indicating that processing is completed and the rack or cartridge can be removed from the loading area when the rack or cartridge should not be accessed. 6. Erroneous ICT Calibration Curves: Erroneous ICT calibration curves might be generated for the Alinity c when there is a malfunction of ICT reference solution delivery, use of wrong/evaporated calibrator, defective ICT modules, etc.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2472-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.