Data Foundry

Medical device recalls 2021

The Azurion series — Class II medical device recall Z-2476-2021

Terminated recall by Philips North America Llc, reported 2021-09-22, distributed nationwide. Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current vo

Recall numberZ-2476-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips North America Llc, Cambridge, MA, United States
Recall initiated2021-08-13
Classified2021-09-14
Reported by FDA2021-09-22
Terminated2024-03-11
DistributedNationwide (US)
Quantity1 units
Reason classesdevice malfunction
Serial numbers921

Product

The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. "Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: "The Azurion series can be used in a hybrid Operation Room. "The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.

Reason for recall

Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2476-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.