Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1 — Class II medical device recall Z-2476-2026
Ongoing recall by GE Medical Systems, LLC, reported 2026-06-24, distributed nationwide. GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end
| Recall number | Z-2476-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Medical Systems, LLC, Waukesha, WI, United States |
| Recall initiated | 2026-05-01 |
| Classified | 2026-06-17 |
| Reported by FDA | 2026-06-24 |
| Distributed | Nationwide (US) |
| Quantity | 23,430 units |
| Reason classes | device malfunction |
| Serial numbers | TAB22260006SA, TAB22260007SA, TAB22260008SA, TAB22260009SA, TAB22340001SA, TAB22360001SA, TAB22360006SA, TAB22360012SA, TAB22360013SA, TAB22360016SA, TAB22360018SA, TAB22370002SA, TAB22370004SA, TAB22370005SA, TAB22370006SA, TAB22370007SA, TAB22370008SA, TAB22370009SA, TAB22370011SA, TAB22370012SA, TAB22370013SA, TAB22370014SA, TAB22370015SA, TAB22370016SA, TAB22370017SA and 475 more |
| Model numbers | 2082844-001-XXX |
Product
Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubator and an infant warmer, Model Numbers 2082844-001-XXX
Reason for recall
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2476-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.