Hitachi Oasis MRI system Product Usage: Hitachi MR system is an… — Class II medical device recall Z-2479-2018
Terminated recall by Hitachi Medical Systems America Inc, reported 2018-07-25, distributed nationwide. The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tablet
| Recall number | Z-2479-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hitachi Medical Systems America Inc, Twinsburg, OH, United States |
| Recall initiated | 2018-06-18 |
| Classified | 2018-07-18 |
| Reported by FDA | 2018-07-25 |
| Terminated | 2020-04-20 |
| Distributed | Nationwide (US) |
| Countries outside the US | MX |
| Quantity | 264 |
Product
Hitachi Oasis MRI system Product Usage: Hitachi MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Reason for recall
The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt.