RefleXion X1 Model RXM1000 UDI Codes: — Class II medical device recall Z-2486-2025

Ongoing recall by Reflexion Medical, Inc., reported 2025-09-10, distributed nationwide. Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour p

Recall numberZ-2486-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmReflexion Medical, Inc., Hayward, CA, United States
Recall initiated2025-07-22
Classified2025-09-02
Reported by FDA2025-09-10
DistributedNationwide (US)
Quantity11
Reason classessoftware
UPC / GTIN / UDI-DI00860003983812
Model numbersRXM1000

Product

RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation therapy system that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage X-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.

Reason for recall

Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).