Breast Implant Pack — Class I medical device recall Z-2492-2014
Terminated recall by Customed, Inc, reported 2014-09-10, distributed nationwide. Customed has determined that there is the possibility that packaging integrity may be compromised on different
| Recall number | Z-2492-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Customed, Inc, Fajardo, PR, United States |
| Recall initiated | 2014-05-20 |
| Classified | 2014-08-29 |
| Reported by FDA | 2014-09-10 |
| Terminated | 2017-08-14 |
| Distributed | Nationwide (US) |
| Countries outside the US | CR, DO, MX, TT |
| Reason classes | sterility, packaging |
| Lot numbers | 112062218, 112093621 |
Product
Breast Implant Pack, Code 900-2660, contains: (1) DRAPE ECONOMY 53 X 77 (3) GOWN SOFT SMS STANDARD XLGE SET IN SLEEVE (3) TOWELS ABSORBENT 15 X 20 (1) MAYO STAND COVER REINFORCED (2) TABLE COVER REINFORCED 50 X 90 (1) SUTURE BAG FLORAL (1) TUBE SUCTION CONNECT Y (6) TOWEL CLOTH HUCK BLUE (4) PAD ABDOMINAL 8 X 7.5 (2) LITE GLOVES (1) CAUTERY TIP POLISHER LIF (1) BLADE SURGICAL CARBON STEEL 15 (1) YANKAUER SUCTION TUBE W/0 VENT (1 0) LAP SPONGE PREWASH 18 X 18 XRD (1) CAUTERY PENCIL ROCKER SWITCH W/HOLSTER (1) UNDERPAD 23 X 36 Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Reason for recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.