MEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S… — Class II medical device recall Z-2492-2026
Ongoing recall by Medline Industries, LP, reported 2026-06-24, distributed nationwide. During an internal review, Medline identified representative samples did not meet applicable performance speci
| Recall number | Z-2492-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medline Industries, LP, Northfield, IL, United States |
| Recall initiated | 2026-05-04 |
| Classified | 2026-06-17 |
| Reported by FDA | 2026-06-24 |
| Distributed | Nationwide (US) |
| Quantity | 1,000 units |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 10889942046126, 40889942046127 |
| Lot numbers | 24BMH490, 24CME465, 24CMF091, 24GMD004, 24IME549, 25CMG042, 25FMJ661, 25JMK709 |
| Model numbers | 22F |
Product
MEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.