RayStation stand-alone software treatment planning system — Class II medical device recall Z-2497-2018

Terminated recall by RAYSEARCH LABORATORIES AB, reported 2018-07-25, distributed nationwide. If the beam model has a highly asymmetric primary source, it is not correctly taken into account in the calcul

Recall numberZ-2497-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRAYSEARCH LABORATORIES AB, Stockholm, Sweden
Recall initiated2018-06-29
Classified2018-07-19
Reported by FDA2018-07-25
Terminated2020-07-16
DistributedNationwide (US)
Quantity746

Product

RayStation stand-alone software treatment planning system, RayStation 4.5, RayStation 4.7, RayStation 4.9 (RayPlan 1), RayStation 5, RayStation 6 (RayPlan 2), RayStation 7 (RayPlan 7) and RayStation 8A (RayPlan 8A) Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Reason for recall

If the beam model has a highly asymmetric primary source, it is not correctly taken into account in the calculation of DMLC fields when the collimator is rotated. This could lead to potentially significant overdosage at delivery. The user must be aware of the issue to avoid incorrect dose calculations during treatment planning.