Philips Azurion System configured with a 1 Phase UPS — Class II medical device recall Z-2502-2025
Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2025-09-10, distributed nationwide. Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the poten
| Recall number | Z-2502-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, N/A, Netherlands |
| Recall initiated | 2025-08-01 |
| Classified | 2025-09-03 |
| Reported by FDA | 2025-09-10 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BE, BG, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DO, EC, EG, ES, FI, FR, GB, GR, HK, HU, IE, IL, IN, IR, IT, JM, JO, JP, KR, LT, LV, MT, MX, NI, NL, NO, NZ, PA, PE, PH, PY, RO, RS, SA, SE, SG, SI, SK, TN, TW, UY, VN, ZA |
| Quantity | 1,861 units |
| UPC / GTIN / UDI-DI | 00884838085251, 00884838085268, 00884838085275, 00884838085282, 00884838085350, 00884838085367, 00884838099203, 00884838099210, 00884838099227, 00884838099234, 00884838099241, 00884838099258, 00884838099265, 00884838099272, 00884838110571 |
| Serial numbers | 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228, 722282 |
| Model numbers | 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228, 722282 |
Product
Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion 3 M15, Model Number 722064. 3. Azurion 7 B12, Model Number 722067. 4. Azurion 7 B20, Model Number 722068. 5. Azurion 7 M12, Model Number 722078. 6. Azurion 7 M20, Model Number 722079. 7. Azurion 3 M12, Model Number 722221. 8. Azurion 3 M15, Model Number 722222. 9. Azurion 7 M12, Model Number 722223. 10. Azurion 7 M20, Model Number 722224. 11. Azurion 7 B12, Model Number 722225. 12. Azurion 7 B20, Model Number 722226. 13. Azurion 5 M12, Model Number 722227. 14. Azurion 5 M20, Model Number 722228. 15. Azurion 7 M20, Model Number 722282.
Reason for recall
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2502-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.