Data Foundry

Medical device recalls 2025

Philips Azurion System configured with a 1 Phase UPS — Class II medical device recall Z-2502-2025

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2025-09-10, distributed nationwide. Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the poten

Recall numberZ-2502-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, N/A, Netherlands
Recall initiated2025-08-01
Classified2025-09-03
Reported by FDA2025-09-10
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BE, BG, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DO, EC, EG, ES, FI, FR, GB, GR, HK, HU, IE, IL, IN, IR, IT, JM, JO, JP, KR, LT, LV, MT, MX, NI, NL, NO, NZ, PA, PE, PH, PY, RO, RS, SA, SE, SG, SI, SK, TN, TW, UY, VN, ZA
Quantity1,861 units
UPC / GTIN / UDI-DI00884838085251, 00884838085268, 00884838085275, 00884838085282, 00884838085350, 00884838085367, 00884838099203, 00884838099210, 00884838099227, 00884838099234, 00884838099241, 00884838099258, 00884838099265, 00884838099272, 00884838110571
Serial numbers722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228, 722282
Model numbers722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228, 722282

Product

Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion 3 M15, Model Number 722064. 3. Azurion 7 B12, Model Number 722067. 4. Azurion 7 B20, Model Number 722068. 5. Azurion 7 M12, Model Number 722078. 6. Azurion 7 M20, Model Number 722079. 7. Azurion 3 M12, Model Number 722221. 8. Azurion 3 M15, Model Number 722222. 9. Azurion 7 M12, Model Number 722223. 10. Azurion 7 M20, Model Number 722224. 11. Azurion 7 B12, Model Number 722225. 12. Azurion 7 B20, Model Number 722226. 13. Azurion 5 M12, Model Number 722227. 14. Azurion 5 M20, Model Number 722228. 15. Azurion 7 M20, Model Number 722282.

Reason for recall

Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2502-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.