NobelBiocare Tri-Lobe-compatible WP X 6 — Class II medical device recall Z-2514-2025

Ongoing recall by Preat Corp, reported 2025-09-10, distributed nationwide. Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relatio

Recall numberZ-2514-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPreat Corp, Santa Maria, CA, United States
Recall initiated2025-06-25
Classified2025-09-03
Reported by FDA2025-09-10
DistributedNationwide (US)
Quantity11
UPC / GTIN / UDI-DI00842092169599
Lot numbers277097
Model numbers9000662

Product

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Reason for recall

Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.