BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106… — Class II medical device recall Z-2514-2026
Ongoing recall by BD SWITZERLAND SARL, reported 2026-06-24, distributed in AK, AL, AR, AZ, CA, CO, DC, FL…. Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread o
| Recall number | Z-2514-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | BD SWITZERLAND SARL, Eysins, Switzerland |
| Recall initiated | 2026-05-21 |
| Classified | 2026-06-18 |
| Reported by FDA | 2026-06-24 |
| Distributed | AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV |
| Countries outside the US | JP |
| Quantity | 2,465,500 units |
| UPC / GTIN / UDI-DI | 00382903949106 |
| Lot numbers | 4277968, 4277973, 4282451, 4282461, 4289801, 4290582, 4292904, 4295848, 4299949, 4303841, 4305108, 4306700, 5129635, 5129642, 5129649, 5129650, 5129651, 5139250, 5139251, 5139252, 5150181, 5153940, 5164061, 5164062, 5178078, 5181442, 5183810, 5183816, 5183824, 5183828, 5218287, 5218291, 5218293, 5218299, 5218302, 5241476, 5247729, 5247731, 5251098, 5275601 and 1 more |
| Model numbers | 394910 |
Product
BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
Reason for recall
Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.