Da Vinci ASSY — Class II medical device recall Z-2521-2026

Ongoing recall by Intuitive Surgical, Inc., reported 2026-07-01, distributed nationwide. Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a s

Recall numberZ-2521-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmIntuitive Surgical, Inc., Sunnyvale, CA, United States
Recall initiated2026-04-22
Classified2026-06-22
Reported by FDA2026-07-01
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, CH, CN, CZ, DE, DK, ES, FR, GB, GR, HK, IE, IL, IT, JP, KR, PL, RO, SA, SE, TR, TW
Quantity454
UPC / GTIN / UDI-DI00886874114605
Serial numbers10003059, 10003509, 10003510, 10008062, 10009706, 10009707, 10012962, 10012963, 10020705, 10020729, 10031144, 10031145, 10035476, 10036960, 10036961, 10043635, 10054478, 10056518, 10065786, 10069899, 10072852, 10072853, 10076490, 10080929, 10087446 and 429 more
Model numbers380601

Product

Da Vinci ASSY, PSS, SP1098, Part Number: 380601

Reason for recall

Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.