Data Foundry

Medical device recalls 2020

uCT 760 System — Class II medical device recall Z-2522-2020

Terminated recall by Shanghai United Imaging Healthcare Co., Ltd., reported 2020-07-15, distributed in TX. Two issues were identified with the computed tomography x-ray system including an intermittent issue which may

Recall numberZ-2522-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmShanghai United Imaging Healthcare Co., Ltd., Shanghai, China
Recall initiated2020-05-08
Classified2020-07-06
Reported by FDA2020-07-15
Terminated2022-01-20
DistributedTX
Quantity5 units
Serial numbers600229, 600230, 600232, 600271, 600282
Model numbers760

Product

uCT 760 System, Computed Tomography X-ray System

Reason for recall

Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2522-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.