Hamilton-T1 — Class I medical device recall Z-2526-2023
Ongoing recall by Hamilton Medical, Inc., reported 2023-09-27, distributed nationwide. Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control
| Recall number | Z-2526-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hamilton Medical, Inc., Reno, NV, United States |
| Recall initiated | 2023-07-26 |
| Classified | 2023-09-20 |
| Reported by FDA | 2023-09-27 |
| Distributed | Nationwide (US) |
| Quantity | 76 units |
| Reason classes | packaging, device malfunction |
| UPC / GTIN / UDI-DI | 07630002801850, 07630002813532 |
| Serial numbers | 15929, 17600, 17661, 17677, 17698, 17699, 17741, 17850, 17877, 17913, 17925, 17955, 17957, 17969, 17975, 17976, 17982, 17985, 17986, 17991, 17993, 17995, 18009, 18166, 18183 and 51 more |
| Model numbers | 161006 |
Product
Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator
Reason for recall
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2526-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.