Multi-Snare Set: 5 mm x 125 cm — Class II medical device recall Z-2526-2026
Ongoing recall by PFM MEDICAL INC., reported 2026-07-01, distributed nationwide. A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular
| Recall number | Z-2526-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | PFM MEDICAL INC., Carlsbad, CA, United States |
| Recall initiated | 2026-04-21 |
| Classified | 2026-06-23 |
| Reported by FDA | 2026-07-01 |
| Distributed | Nationwide (US) |
| Quantity | 32 |
| Reason classes | foreign material |
| UPC / GTIN / UDI-DI | 04042301033567, 04042301061829 |
| Lot numbers | 125, 147305, 147310 |
| Model numbers | 147305, 147310 |
Product
Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
Reason for recall
A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2526-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.