Philips M3015A Microstream CO2 Extension Model No — Class II medical device recall Z-2531-2020
Terminated recall by Philips North America, LLC, reported 2020-07-15, distributed nationwide. Affected M3015A extensions are incompatible with the lntelliVue X3 Patient Monitor (Model No. 867030) and lnte
| Recall number | Z-2531-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips North America, LLC, Andover, MA, United States |
| Recall initiated | 2020-06-08 |
| Classified | 2020-07-07 |
| Reported by FDA | 2020-07-15 |
| Terminated | 2023-01-09 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EE, EG, ES, FI, FR, GB, GH, GR, GT, HK, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MX, MY, NI, NL, NO, NZ, OM, PE, PH, PL, PT, PY, QA, SA, SE, SG, SI, SK, TH, TR, TW, UA, ZA |
| Quantity | 46,597 units |
| Model numbers | 862393, 867030, 867036 |
Product
Philips M3015A Microstream CO2 Extension Model No. 862393 -used with an lntelliVue X3 Patient Monitor (Model No. 867030) or a lntelliVue MMX Multi-Measurement Module (Model No. 867036)
Reason for recall
Affected M3015A extensions are incompatible with the lntelliVue X3 Patient Monitor (Model No. 867030) and lntelliVue MMX multi-measurement server (Model No. 867036).If they are connected to a host monitor, communication with the X3 or MMX could be disrupted, resulting in a loss of monitoring on the host monitor.