VITROS 5 — Class II medical device recall Z-2544-2015
Terminated recall by Ortho-Clinical Diagnostics, reported 2015-09-09, distributed nationwide. Software anomaly may occur during the installation of an Assay Data Diskette (ADD) using Software Versions 2.2
| Recall number | Z-2544-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Rochester, NY, United States |
| Recall initiated | 2015-06-30 |
| Classified | 2015-09-02 |
| Reported by FDA | 2015-09-09 |
| Terminated | 2018-05-29 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE |
| Reason classes | potency, software |
| UPC / GTIN / UDI-DI | 10758750001132, 10758750001644 |
| Serial numbers | 34000119, J34000133, J34000193, J34000197, J34000212, J34000285, J34000316, J34000318, J34000337, J34000360, J34000420, J34000451, J34000478, J34000495, J34000497, J34000529, J34000538, J34000552, J34000557, J34000590, J34000596, J34000601, J34000631, J34000642, J34000761 and 39 more |
| Model numbers | 6801375, 6801890 |
Product
VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique Device Identifier No. 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number/Product Code 6801890, Unique Device Identifier No. 10758750001644, IVD --- Ortho Clinical Diagnostics Inc. The VITROS 5,1 FS Chemistry System with enGEN Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)
Reason for recall
Software anomaly may occur during the installation of an Assay Data Diskette (ADD) using Software Versions 2.2.1 through 2.8. If the user is utilizing User- Configured (Modified) parameters, ALL default settings are restored in the VITROS 5,1 FS Chemistry System from the ADD instead of retaining the user modified parameters.