Hillrom VOLARA System P — Class I medical device recall Z-2551-2026
Ongoing recall by Baxter Healthcare Corporation, reported 2026-07-08, distributed nationwide. There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system p
| Recall number | Z-2551-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Baxter Healthcare Corporation, Deerfield, IL, United States |
| Recall initiated | 2026-05-21 |
| Classified | 2026-07-01 |
| Reported by FDA | 2026-07-08 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA |
| Quantity | 13,881 units |
| Reason classes | packaging |
| UPC / GTIN / UDI-DI | 10887761981499 |
| Model numbers | M08474 |
Product
Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason for recall
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2551-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.