CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor — Class II medical device recall Z-2554-2014
Terminated recall by Curbell Medical, Inc., reported 2014-09-03, distributed nationwide. The firm became aware of a potential problem that was initiated by a customer complaint. After consultation wi
| Recall number | Z-2554-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Curbell Medical, Inc., Orchard Park, NY, United States |
| Recall initiated | 2014-08-05 |
| Classified | 2014-08-26 |
| Reported by FDA | 2014-09-03 |
| Terminated | 2017-02-28 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, IE, IT |
| Lot numbers | 0414 |
| Model numbers | CSM-BC400 |
Product
CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor, MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor; CLASSIFICATION NAME: Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm.
Reason for recall
The firm became aware of a potential problem that was initiated by a customer complaint. After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor. This change to the resistor was a planned change to address a product improvement (improve battery drain). However, this modification rendered the monitor's alarm via the nurse call communication port to not function.