cobas m 511 integrated hematology analyzer — Class II medical device recall Z-2557-2018

Terminated recall by Roche Diagnostics Hematology, reported 2018-08-01. Discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalasse

Recall numberZ-2557-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Hematology, Westborough, MA, United States
Recall initiated2018-05-07
Classified2018-07-26
Reported by FDA2018-08-01
Terminated2020-05-12
Countries outside the USAT, CH, DE, FR, HK, JP, KR, NL, PK, SG, TH, VN
Quantity17
Model numbers07261691190

Product

cobas m 511 integrated hematology analyzer, Model Number 07261691190 Product Usage: The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET.

Reason for recall

Discrepant results have been reported in patients with severe microcytic anemia (e.g.iron deficiency, thalassemia) and the following parameters are affected: RBC (red blood cell count), HGB (hemoglobin concentration), MCH (mean corpuscular hemoglobin), HCT (hematocrit), and MCV (mean corpuscular volume).