Calibrate — Class II medical device recall Z-2557-2023
Completed recall by Landauer, reported 2023-09-20. LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +
| Recall number | Z-2557-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Completed |
| Voluntary or mandated | Voluntary |
| Recalling firm | Landauer, Glenwood, IL, United States |
| Recall initiated | 2023-07-12 |
| Classified | 2023-09-12 |
| Reported by FDA | 2023-09-20 |
| Quantity | 5,958 units |
| UPC / GTIN / UDI-DI | 00860003399903 |
Product
Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 04217-000 (EMEA number VINLNAN003); b) 04217-KIT; c) 04217-SET; radiation monitoring dosimeter used with the microSTAR readers
Reason for recall
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.