QC — Class II medical device recall Z-2558-2023

Completed recall by Landauer, reported 2023-09-20. LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +

Recall numberZ-2558-2023
Product typeMedical device
ClassificationClass II
StatusCompleted
Voluntary or mandatedVoluntary
Recalling firmLandauer, Glenwood, IL, United States
Recall initiated2023-07-12
Classified2023-09-12
Reported by FDA2023-09-20
Quantity2,240 units
UPC / GTIN / UDI-DI00860003399903

Product

QC (80 kVp) nanoDot D2DNS, Model Numbers: a) 04218-000 (EMEA number VINLNAN004); b) 04218-KIT; c) 04218-SET; radiation monitoring dosimeter used with the microSTAR readers

Reason for recall

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.