Data Foundry

Medical device recalls 2014

PERSONA The Personalized Knee System Constrained Tibial Articular… — Class II medical device recall Z-2575-2014

Terminated recall by Zimmer, Inc., reported 2014-09-10, distributed nationwide. Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee

Recall numberZ-2575-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer, Inc., Warsaw, IN, United States
Recall initiated2014-08-07
Classified2014-09-04
Reported by FDA2014-09-10
Terminated2015-01-27
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, CH, DE, ES, FR, GB, IL, IN, IT, JP, KR, LU, NL, SA, SG, ZA
Quantity77,737 units
Reason classesdevice malfunction
Lot numbers62345370, 62347233, 62347240, 62351310, 62352330, 62357328, 62357329, 62357865, 62357866, 62358632, 62358633, 62363685, 62364505, 62372411, 62373739, 62373740, 62374803, 62399612, 62399870, 62402721, 62405896, 62406970, 62407568, 62407772, 62410788, 62410789, 62410790, 62412999, 62416376, 62423921, 62423922, 62424796, 62424804, 62428432, 62430403, 62432150, 62432152, 62432154, 62432157, 62432790 and 63 more
Model numbers42517600303, 42517600313, 42517600505, 42517600515, 42517600707, 42517600717, 42517600909, 42517600919, 42527600303, 42527600313, 42527600505, 42527600515, 42527600707, 42527600717, 42527600909, 42527600919

Product

PERSONA The Personalized Knee System Constrained Tibial Articular Surface Provisional BOTTOM with the following: L CD +0MM; L CD +6MM; L EF +0MM;L EF +6MM; L GH +0MM; L GH +6MM; L J +0MM; L J +6MM; R CD +0MM; R CD +6MM; R EF +0MM; R EF +6MM; R GH +0MM; R GH +6MM; R J +0MM; R J +6MM; and

Reason for recall

Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2575-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.