BeamAdjust software Version 2 — Class II medical device recall Z-2577-2023

Ongoing recall by PTW-FREIBURG, reported 2023-09-20, distributed nationwide. Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000

Recall numberZ-2577-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPTW-FREIBURG, Freiburg Im Breisgau, Germany
Recall initiated2023-06-07
Classified2023-09-11
Reported by FDA2023-09-20
DistributedNationwide (US)
Quantity42 us
Model numbersS080032

Product

BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320 ): " OCTAVIUS Detector 1600 SRS (T10056) (UDI-DI: EPTWT100560) " OCTAVIUS Detector 1600 MR (T10057) (UDI-DI: EPTWT100570) " OCTAVIUS Detector 1600 XDR (T10058) (UDI-DI: EPTWT100580) " OCTAVIUS Detector 1000 SRS (T10036) (UDI-DI: EPTWT100360)

Reason for recall

Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.