BeamAdjust software Version 2 — Class II medical device recall Z-2577-2023
Ongoing recall by PTW-FREIBURG, reported 2023-09-20, distributed nationwide. Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000
| Recall number | Z-2577-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | PTW-FREIBURG, Freiburg Im Breisgau, Germany |
| Recall initiated | 2023-06-07 |
| Classified | 2023-09-11 |
| Reported by FDA | 2023-09-20 |
| Distributed | Nationwide (US) |
| Quantity | 42 us |
| Model numbers | S080032 |
Product
BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320 ): " OCTAVIUS Detector 1600 SRS (T10056) (UDI-DI: EPTWT100560) " OCTAVIUS Detector 1600 MR (T10057) (UDI-DI: EPTWT100570) " OCTAVIUS Detector 1600 XDR (T10058) (UDI-DI: EPTWT100580) " OCTAVIUS Detector 1000 SRS (T10036) (UDI-DI: EPTWT100360)
Reason for recall
Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.