PORTEX Bivona Tracheostomy Silicone Tube — Class I medical device recall Z-2592-2024
Ongoing recall by Smiths Medical ASD Inc., reported 2024-09-04. Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and
| Recall number | Z-2592-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Smiths Medical ASD Inc., Minneapolis, MN, United States |
| Recall initiated | 2024-05-29 |
| Classified | 2024-08-23 |
| Reported by FDA | 2024-09-04 |
| UPC / GTIN / UDI-DI | 15021312014716, 15021312014723, 15021312014730, 15021312014747 |
| Serial numbers | 3840062, 3840087, 3853439, 3868231, 3874300, 3877647, 3877660, 3880697, 3884269, 3891522, 3894865, 3894895, 3901301, 3904684, 3907972, 3917986, 3917996, 3920949, 3927831, 3930057, 3931151, 3939632, 3939639, 3939659, 3942500 and 105 more |
| Model numbers | 60NFP25, 60NFP30, 60NFP35, 60NFP40 |
Product
PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5FLEXTEND PLUS , Product Code/List Number/Item Code 60NFP25; b) NEONATAL TRACHEOSTOMY TUBE 3.0FLEXTEND PLUS , Product Code/List Number/Item Code 60NFP30; c) NEONATAL TRACHEOSTOMY TUBE 3.5FLEXTEND PLUS , Product Code/List Number/Item Code 60NFP35; d) NEONATAL TRACHEOSTOMY TUBE 4.0FLEXTEND PLUS , Product Code/List Number/Item Code 60NFP40
Reason for recall
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2592-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.