Hospira RECEPTAL CANISTERS: 1L — Class II medical device recall Z-2593-2016
Terminated recall by Hospira Inc., reported 2016-08-24, distributed nationwide. Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters,
| Recall number | Z-2593-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hospira Inc., Lake Forest, IL, United States |
| Recall initiated | 2013-09-10 |
| Classified | 2016-08-18 |
| Reported by FDA | 2016-08-24 |
| Terminated | 2016-12-15 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, CR, SG, ZA |
Product
Hospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL) Canister, ATS, 2000, with Valve 1) List Numbers 43449-01 and 43449-11; LIST NO. 43449 RECEPTAL CANISTER; CANISTER FOR REUSE, THROW AWAY LINER ONLY; DO NOT FILL 900 ABOVE LINE 2) List Numbers 43423-01 and 43423-21; LIST NO. 43423; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; 1400. 3) List Numbers 43445-01 and 43445-11; LIST NO. 43445; RECEPTAL CANISTER; CANISTER FOR REUSE; THROW AWAY LINER ONLY; DO NOT FILL 1900 ABOVE THIS LINE. 4) List Number 43496-01.
Reason for recall
Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.
Machine access