Data Foundry

Medical device recalls 2021

Flow sensors kept as stand-alone user replaceable spare parts for the… — Class II medical device recall Z-2595-2021

Ongoing recall by GE Healthcare, LLC, reported 2021-10-06, distributed nationwide. The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cau

Recall numberZ-2595-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2021-08-02
Classified2021-09-30
Reported by FDA2021-10-06
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BM, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA
Quantity310,072 flow
Reason classespackaging

Product

Flow sensors kept as stand-alone user replaceable spare parts for the GEHC anethesia machines: 2089610-001 FLOW SENSOR, LEGACY VAR ORF BCG (blue, cleanable) 2089610-001-S FLOW SENSOR, LEGACY VAR ORF BCG, SERVICE (blue, cleanable) 2087640-001 FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG (gray, autoclavable) 2087640-001-S FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG, SERVICE (gray, autoclavable) 2096513-001-S FLOW SENSOR ASSEMBLY 5697309 R-FMI34109-FLOW SENSOR, LEGACY VAR ORF BCG 5697310 R-FMI34109-FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG

Reason for recall

The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of volume to the patient.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2595-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.