Flow sensors kept as stand-alone user replaceable spare parts for the… — Class II medical device recall Z-2595-2021
Ongoing recall by GE Healthcare, LLC, reported 2021-10-06, distributed nationwide. The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cau
| Recall number | Z-2595-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, LLC, Waukesha, WI, United States |
| Recall initiated | 2021-08-02 |
| Classified | 2021-09-30 |
| Reported by FDA | 2021-10-06 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BG, BH, BM, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA |
| Quantity | 310,072 flow |
| Reason classes | packaging |
Product
Flow sensors kept as stand-alone user replaceable spare parts for the GEHC anethesia machines: 2089610-001 FLOW SENSOR, LEGACY VAR ORF BCG (blue, cleanable) 2089610-001-S FLOW SENSOR, LEGACY VAR ORF BCG, SERVICE (blue, cleanable) 2087640-001 FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG (gray, autoclavable) 2087640-001-S FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG, SERVICE (gray, autoclavable) 2096513-001-S FLOW SENSOR ASSEMBLY 5697309 R-FMI34109-FLOW SENSOR, LEGACY VAR ORF BCG 5697310 R-FMI34109-FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG
Reason for recall
The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of volume to the patient.