MEDLINE STERILE IRIS SCISSORS CVD/STD — Class II medical device recall Z-2595-2023

Ongoing recall by MEDLINE INDUSTRIES, LP - Northfield, reported 2023-09-27, distributed nationwide. 4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector t

Recall numberZ-2595-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMEDLINE INDUSTRIES, LP - Northfield, Northfield, IL, United States
Recall initiated2023-08-03
Classified2023-09-18
Reported by FDA2023-09-27
DistributedNationwide (US)
Countries outside the USCA, HK, SG, TW
Quantity324,451 devices
Reason classessterility
UPC / GTIN / UDI-DI10080196847273
Model numbersDYNJ04049

Product

MEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049

Reason for recall

4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2595-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.