Flow sensors that may be installed in the following GEHC anesthesia… — Class II medical device recall Z-2596-2021

Ongoing recall by GE Healthcare, LLC, reported 2021-10-06, distributed nationwide. The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cau

Recall numberZ-2596-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2021-08-02
Classified2021-09-30
Reported by FDA2021-10-06
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BM, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA
Quantity310,072 flow
Reason classespackaging
UPC / GTIN / UDI-DI00080682102339, 00840682102292, 00840682102322

Product

Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3) Aestiva 7100; (4) Aespire System Carestation, S/5 Protiva Config., Aespire 100 Compact Wall Mount (Aespire 7100); (5) S/5 Aespire 7900; (6) Aespire View; (7) Avance; (8) Avance CS2, GTIN 00840682102292; (9) Aisys; (10) Aisys CS2, GTIN 00840682102322; (11) Amingo; (12) 9100 NXT Config, Aelite NXT; (13) Protiva 7100;

Reason for recall

The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of volume to the patient.