Data Foundry

Medical device recalls 2026

Medline Convenience kits containing recalled Swan-Ganz Catheters… — Class II medical device recall Z-2608-2026

Ongoing recall by Medline Industries, LP, reported 2026-07-08, distributed nationwide. Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breaka

Recall numberZ-2608-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedline Industries, LP, Northfield, IL, United States
Recall initiated2026-05-28
Classified2026-06-26
Reported by FDA2026-07-08
DistributedNationwide (US)
Quantity9,272 kits
Reason classespackaging, device malfunction
UPC / GTIN / UDI-DI10198459704543
Lot numbers26DBB599

Product

Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407

Reason for recall

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.